Light Source
|
The life for the
convergent tungsten lamp is 5000 hours.
|
The outer
light source uses transmitting deuterium lamp. The deuterium lamp and
tungsten lamp can be controlled separately. (not standard)
|
Measurement Mode
|
Special test module for water quality
test
Spectrum scanning and peak-picking.
Quantitative measurement
Kinetics
Photometric measurement
|
Monochromator
|
Concave plate holographic grating
|
Detector
|
PORS16: CCD S8378,1024 Pixels / PORS16V:CCD ILX511 2048 Pixels
|
Measurement parts
|
Immerging fiber detector, 10mm pathlength 20mm Pathlength (Exchange)
|
Power supply
|
Inner chargeable battery and one charging
for one-day work.
|
With vehicle-loading power supply adaptor
(not standard)
|
Charger
|
Operating system
|
Embedded operation system, integration
2M-Flash memorizer. With timing function.
|
Input/display
|
Large touch-screen, 320×240 lattice with backlight
|
Specifications
|
Wavelength range
|
380nm-800nm / 200nm –
800nm
(standard, the
main unit has a tungsten lamp)
|
Data resolution
|
0.6nm
|
Spectral bandwidth
|
4±1nm
|
Wavelength accuracy
|
<1.0 nm
|
Wavelength repeatability
|
0.1nm
|
Baseline flatness
|
±0.005Abs
|
Noise
|
±0.003Abs
|
Integral time
|
0.005s-25s
|
Scanning speed
|
>4200nm/s
|
Measurement precision
|
±2%
|
Measurement item
|
Measurement Accuracy
|
Measurement range
|
COD(low concentration)
|
<±10%
|
10mg/L-150 mg/L
|
COD(high concentration)
|
<±10%
|
150mg/L-1500 mg/L
|
Ammonia nitrogen
(Nesslers reagent colorimetric method)
|
<±10%
|
0.03mg/L-2.5 mg/L
|
Cr 6+
|
<±8%
|
0.01mg/L-0.6 mg/L
|
Cyanide
|
<±8%
|
0.001mg/L-0.2 mg/L
|
CODMn
|
<±15%
|
1.0mg/L-10.0 mg/L
|
Ammonia (hypo-bromate
oxidation method)
|
<±15%
|
0.01mg/L-0.2 mg/L
|
Nitrite
|
<±8%
|
0.01mg/L-0.2 mg/L
|
|
Nitrate(Cd deoxidization
method)
|
\
|
0.02mg/L-0.5 mg/L
|
Inorganic P(P-Mo method)
|
<±10%
|
0.05mg/L-1.0 mg/L
|
Dimension
|
267×170×88mm
|
Environmental temperature
|
Operation environment 0-40℃
Store environment 0-50℃
|